First experience of the usage of a GORE CARDIOFORM Septal Occluder device for treatment of a significant residual commissural mitral regurgitation jet following a MitraClip procedure.
Yoshifumi NakajimaSaibal KarPublished in: Catheterization and cardiovascular interventions : official journal of the Society for Cardiac Angiography & Interventions (2017)
An 81-year-old male presented with symptomatic recurrent MR following a MitraClip procedure using two clips. After adding another clip, there was still a residual jet medially. We plugged the gap using 12-mm AMPLATZER Muscular VSD Occluder device (St. Jude Medical, Minneapolis, MN). MR was reduced but hemolysis occurred. We retrieved the VSD Occluder device to improve hemolysis, but the MR jet worsened to severe, so we placed a 30-mm GORE CARDIOFORM Septal Occluder device. The device was stable, the MR jet was reduced dramatically, and hemolysis was not observed. Materials of these devices are mesh fabric made from nitinol and polyester in a VSD Occluder device and expanded polytetrafluoroethylene in GORE CARDIOFORM Septal Occluder device. The difference between device materials affects flow pattern of residual jet and hemolysis. In conclusion, the patient's risk for hemolysis is potentially affected by the occluder or plug material. GORE CARDIOFORM is a feasible choice to reduce the risk of hemolysis in a selected patient.