Login / Signup

Sequential Two-Photon Delayed Fluorescence Anisotropy for Macromolecular Size Determination.

Yi-Han LuMatthew C JenkinsKatherine G RichardsonSayan PaluiMd Shariful IslamJagnyaseni TripathyNeal K DevarajRobert M Dickson
Published in: The journal of physical chemistry. B (2023)
Time-resolved fluorescence anisotropy (FA) uses the fluorophore depolarization rate to report on rotational diffusion, conformation changes, and intermolecular interactions in solution. Although FA is a rapid, sensitive, and nondestructive tool for biomolecular interaction studies, the short (∼ns) fluorescence lifetime of typical dyes largely prevents the application of FA on larger macromolecular species and complexes. By using triplet shelving and recovery of optical excitation, we introduce optically activated delayed fluorescence anisotropy (OADFA) measurements using sequential two-photon excitation, effectively stretching fluorescence anisotropy measurement times from the nanosecond scale to hundreds of microseconds. We demonstrate this scheme for measuring slow depolarization processes of large macromolecular complexes, derive a quantitative rate model, and perform Monte Carlo simulations to describe the depolarization process of OADFA at the molecular level. This setup has great potential to enable future biomacromolecular and colloidal studies.
Keyphrases
  • energy transfer
  • monte carlo
  • single molecule
  • quantum dots
  • living cells
  • high resolution
  • molecular dynamics
  • case control
  • high speed
  • current status
  • zika virus
  • mouse model