Performance Assessment of Sysmex DI-60: Is Digital Morphology Analyzer Reliable for White Blood Cell Differentials in Body Fluids?

Eunju ShinMina HurHanah KimGun-Hyuk LeeMi-Hyun HongMinjeong NamSeungho Lee
Published in: Diagnostics (Basel, Switzerland) (2024)
The DI-60 may provide reliable data in neutrophil- and lymphocyte-dominant BF samples. However, it may require longer times and higher workloads for WBC differentials especially in BF samples containing atypical cells. Further improvement would be needed before applying DM analyzers for routine clinical practice in BF analysis.