Ocular biometry through fully refocused steady-state magnetic resonance imaging sequence: reliability and agreement with the IOLMaster® 500.
Lorenzo Ismael Pérez-SánchezJulia Gutierrez-VazquezMaria Satrustegui-LapetraFrancisco Ferreira-ManuelJuan Jose Arevalo-MansoJuan Jesus Gomez-HerreraJuan-José Criado-ÁlvarezPublished in: International ophthalmology (2021)
ACD and AL measurements using FRSS sequencing present high LoA and reliability when compared with partial coherence interferometry using the IOLMaster® 500. The results were better for FRSS in pseudo-color scale in ACD determination and for FRSS in greyscale in AL determination. FRSS would not be recommended for IOL power calculation due to variability of AL measurement.