The impact of a prophylactic skin dressing on surface-guided patient positioning in chest wall Radiation Therapy.

James CummingKenton ThompsonKatrina WoodfordVanessa PanettieriDaniel Sapkaroski
Published in: Journal of medical radiation sciences (2024)
Mepitel can impact the accuracy of SGRT patient-positioning in chest wall RT. The variation however is small and unlikely to have any clinical impact if SGRT is coupled with image guidance and appropriate PTV margins. Further investigation is required to assess the effect of Mepitel on SGRT accuracy in other treatment sites, as well as any potential dosimetric impacts.