Fracture Resistance of Zirconia Abutments with or without a Titanium Base: An In Vitro Study for Tapered Conical Connection Implants.

Shota WatanabeTamaki NakanoShinji OnoYasufumi YamanishiTakashi MatsuokaShoichi Isigaki
Published in: Materials (Basel, Switzerland) (2022)
Dental implants with tapered conical connections are often combined with zirconia abutments for esthetics; however, the effect of the titanium base on the implant components remains unclear. This study evaluated the effects of a titanium base on the fracture resistance of zirconia abutments and damage to the tapered conical connection implants. Zirconia (Z) and titanium base zirconia (ZT) abutments were fastened to Nobel Biocare (NB) implants and Straumann (ST) implants and subjected to static load testing according to ISO 14801:2016. The experiments were performed with 3 mm of the platform exposed (P3) and no platform exposed (P0). The fracture loads were statistically greater in the titanium base abutments than the zirconia abutments for the NB and ST specimens in the P0 condition. In the P3 condition of the ST specimens, the deformation volume of the ZT group was significantly greater than the Z group. The titanium base increased the fracture resistance of the zirconia abutments. Additionally, the titanium base caused more deformation in the P3 condition. The implant joint design may also affect the amount of damage to the implants when under a load. The mechanical properties of the abutment should be considered when selecting a clinical design.