Comparative Performance Analysis of Urised 3 and DIRUI FUS-200 Automated Urine Sediment Analyzers and Manual Microscopic Method.
Emre YalcinkayaHayriye ErmanEray KiracAfife SerifogluAlperen AksoyFerruh Kemal IsmanMustafa B CekmenPublished in: Medeniyet medical journal (2019)
The concordance between the analyzers and manual microscopy for WBC was better than that of RBC. The concordance between the two analyzers was better for WBC and RBC, with respect to the manual microscopy. Although the Urised 3, FUS-200 and manual microscopy counts were in agreement; confirmation of the results of automated analyzers with manual microscopy is particularly helpful, for pathological samples with near cut-off values.