Speech recognition, loudness, and preference with extended bandwidth hearing aids for adult hearing aid users.

Maaike Van EeckhouttePaula FolkeardDanielle GlistaSusan Scollie
Published in: International journal of audiology (2020)
Objective: In contrast to the past, some current hearing aids can provide gain for frequencies above 4-5 kHz. This study assessed the effect of wider bandwidth on outcome measures using hearing aids fitted with the DSL v5.0 prescription.Design: There were two conditions: an extended bandwidth condition, for which the maximum available bandwidth was provided, and a restricted bandwidth condition, in which gain was reduced for frequencies above 4.5 kHz. Outcome measures were assessed in both conditions.Study sample: Twenty-four participants with mild-to-moderately-severe sensorineural high-frequency sloping hearing loss.Results: Providing extended bandwidth resulted in maximum audible output frequency values of 7.5 kHz on average for an input level of 65 dB SPL. An improvement in consonant discrimination scores (4.1%), attributable to better perception of /s/, /z/, and /t/ phonemes, was found in the extended bandwidth condition, but no significant change in loudness perception or preferred listening levels was found. Most listeners (79%) had either no preference (33%) or some preference for the extended bandwidth condition (46%).Conclusions: The results suggest that providing the maximum bandwidth available with modern hearing aids fitted with DSL v5.0, using targets from 0.25 to 8 kHz, can be beneficial for the tested population.