Login / Signup

Sacroiliac joint stabilization using implants provide better fixation in females compared to males: a finite element analysis.

Amin JoukarAli KiapourAnoli ShahArdalan Seyed VosoughiVijay K Goel
Published in: European spine journal : official publication of the European Spine Society, the European Spinal Deformity Society, and the European Section of the Cervical Spine Research Society (2021)
Implants exhibited higher stresses in the female model compared to the male model. However, chances of SIJ implant failure in the female patients are still minimal, based on the calculated factor of safety which is still very high. Both lateral and posterior surgical approaches were effective in both sexes; however, the lateral approach may provide a better biomechanical response, especially for females. Moreover, implant design characteristics did not make a difference in the implants' biomechanical performance. SIJ stabilization was primarily provided by the implants which were the farthest from the sacrum rotation center.
Keyphrases