A multi-institutional evaluation of small field output factor determination following the recommendations of IAEA/AAPM TRS-483.

Wolfgang LechnerRodolfo AlfonsoMehenna AribM Saiful HuqAnas IsmailRajesh KinhikarJosé M Lárraga-GutiérrezKarthick Raj ManiNkosingiphile MaphumuloOtto A SauerShaima ShoeirSivalee SuriyapeeKaren Christaki
Published in: Medical physics (2022)
Quality assurance on the SFOF measurements can be done by investigation of the spread of SFOFs measured with multiple detectors and by comparison to linac-type specific SFOFs. A follow-up of a measurement session should be conducted if the spread of SFOFs is larger than 5%, 3%, and 2% for field sizes of 0.5 cm × 0.5 cm, 1 cm × 1 cm, and field sizes larger than 2 cm × 2 cm, respectively. Additionally, deviations of measured SFOFs to the linac-type-curves of more than 7%, 3%, and 2% for field sizes 0.5 cm × 0.5 cm, 1 cm × 1 cm, and field sizes larger than 1 cm × 1 cm, respectively, should be followed up.