Versatility and stability optimization of flow-focusing droplet generators via quality metric-driven design automation.

David P McIntyreAli LashkaripourDiana ArguijoPolly Morrell FordyceDouglas M Densmore
Published in: Lab on a chip (2023)
Droplet generation is a fundamental component of droplet microfluidics, compartmentalizing biological or chemical systems within a water-in-oil emulsion. As adoption of droplet microfluidics expands beyond expert labs or integrated devices, quality metrics are needed to contextualize the performance capabilities, improving the reproducibility and efficiency of operation. Here, we present two quality metrics for droplet generation: performance versatility, the operating range of a single device, and stability, the distance of a single operating point from a regime change. Both metrics were characterized in silico and validated experimentally using machine learning and rapid prototyping. These metrics were integrated into a design automation workflow, DAFD 2.0, which provides users with droplet generators of a desired performance that are versatile or flow stable. Versatile droplet generators with stable operating points accelerate the development of sophisticated devices by facilitating integration of other microfluidic components and improving the accuracy of design automation tools.
Keyphrases
  • single cell
  • high throughput
  • quality improvement
  • circulating tumor cells