Login / Signup

Comparison of 2 modern swept-source optical biometers-IOLMaster 700 and Anterion.

Achim LangenbucherNóra SzentmáryAlan CaylessJascha WendelsteinPeter Hoffmann
Published in: Graefe's archive for clinical and experimental ophthalmology = Albrecht von Graefes Archiv fur klinische und experimentelle Ophthalmologie (2022)
The biometric measures used in standard formulae (AL, corneal front surface curvature/power) are consistent between instruments. However, modern formulae involving ACD, CCT or corneal back surface curvature may yield differences in IOLP, and therefore, formula constant optimisation customised to the biometer type is required.
Keyphrases
  • optical coherence tomography
  • wound healing
  • high resolution
  • cataract surgery
  • high speed
  • mass spectrometry
  • preterm infants