Sensitivity and specificity of SS-OCT for detecting macular pathologies vs SD-OCT.
Elizabeth YeuJohn P BerdahlPreeya K GuptaMichael PattersonPublished in: Journal of cataract and refractive surgery (2024)
Using retinal SS-OCT biometer scans as a replacement of the dedicated macular SD-OCT for screening or diagnosing macular health would not be appropriate because of its low sensitivity. SS-OCT biometer may potentially fail to identify approximately one-fourth of patients who actually have the disease. Therefore, the final decision on macular health should be based on the gold standard SD-OCT scans. When full macular SD-OCT scans are not accessible, the limited retinal scan information from SS-OCT biometer may still provide useful insights into the macular health.