Accuracy of Intraocular Lens Power Calculation Using Anterior Chamber Depth from Two Devices with Barrett Universal II Formula.
Hannah Muniz CastroAudrey X TaiSamuel J SampsonMatthew WadeMarjan FaridSumit GargPublished in: Journal of ophthalmology (2019)
The Catalys biometry yielded a significantly larger ACD value than the IOLMaster. This difference in ACD value, however, did not reflect in a statistically significant difference in IOL power calculation and refractive prediction error using the Barrett Universal II Formula.