Login / Signup

Pilot investigation into osteotome hard surface coating and cutting-edge degradation.

David Edward WhiteJim BartleyChristopher WhittingtonLorenzo M G GarciaKaushik ChandCeline Turangi
Published in: Journal of orthopaedic surgery and research (2020)
The rapid loss of blade sharpness found in the as-supplied osteotome supports the degradation in cutting performance frequently reported by surgeons. The findings from this study demonstrate blade sharpness index better detects cutting-edge wear compared to visual inspection. Results from this pilot study also suggest the coating of osteotomes in hard-wearing biocompatible materials assists in retaining cutting-edge sharpness over multiple uses. Further study using a larger sample size is required to validate these findings.
Keyphrases
  • quality improvement
  • study protocol
  • clinical trial
  • drug release
  • double blind