Login / Signup

Finite Element Analysis of Dental Implants with Zirconia Crown Restorations: Conventional Cement-Retained vs. Cementless Screw-Retained.

Jae Hyun LeeHo Yeol JangSu Young Lee
Published in: Materials (Basel, Switzerland) (2021)
The present study was designed to compare the stress distributions in two restoration types of implants and the surrounding bone. The first restoration type was a conventional cement-retained zirconia crown, and the second was a novel cementless screw-retained zirconia crown with a base abutment. A three-dimensional finite element method was used to model the implants, restorations, and supporting bone. A comparative study of the two implants was performed under two masticatory loads: a vertical load of 100 N and a 30-degree oblique load of 100 N. Under both loading conditions, the maximum von Mises stress and strain values in the implant and supporting bone were higher in the conventional cement-retained restoration model than in the cementless screw-retained model. In terms of stress distribution, the cementless screw-retained zirconia crown with base abutment may be considered a superior restoration option compared to the conventional cement-retained zirconia crown.
Keyphrases
  • total knee arthroplasty
  • soft tissue
  • finite element analysis
  • total hip arthroplasty
  • total hip
  • bone mineral density
  • finite element
  • bone loss