Login / Signup

Effect of Plate Length on Construct Stiffness and Strain in a Synthetic Short-Fragment Fracture Gap Model Stabilized with a 3.5-mm Locking Compression Plate.

Fabian N TrefnyMark GlydeGiselle HosgoodAlex HayesRobert Day
Published in: Veterinary and comparative orthopaedics and traumatology : V.C.O.T (2024)
 Incremental increases in construct stiffness and incremental decreases in plate strain were only identified when comparing 6-hole, 8-hole, and 10-hole plates to 12-hole plates, and 6-hole to 12-hole plates, respectively. Strain on the bone model showed an incremental decrease when comparing 6-hole to 10-hole and 12-hole plates. A long plate offered biomechanical advantages of increased construct stiffness and reduced plate and bone model strain, over a short plate in this in vitro model.
Keyphrases
  • perovskite solar cells
  • solar cells
  • bone mineral density
  • body composition
  • bone regeneration