Biomechanical analysis of push-pull principle versus traditional approach in locking plates for proximal humeral fracture treatment.

Guy PutzeysEdoardo BoriTom OveresBernardo Innocenti
Published in: Archives of orthopaedic and trauma surgery (2024)
More homogeneous stress distribution is found with the push-pull plate in all three testing set-ups, showing lower unloaded areas (and thus lower stress-shielding) compared to the traditional plate; the screws implemented returned to be all loaded in at least one of the set-ups, thus showing that they all contribute to plate stability.