Proposal for a New Fixation Method for Pauwels Type III Femoral Neck Fracture-Metaphyseal Stem: A Finite-Element Analysis.

Anderson FreitasFabrício Reichert BarinLeonardo Rigobello BattaglionHélio Ismael da CostaEduardo Duarte SantosHugo Amando RosadoMarcos GiordanoVincenzo GiordanoAntonio Carlos Shimano
Published in: Indian journal of orthopaedics (2021)
The present FEM study showed that the MNS device increases general stresses and reduces localized stresses, when compared to the DHS + ARS and ASNIS constructions used to fix Pauwels type-III femoral neck fracture in synthetic models. In this sense, the MNS showed a low fracture focus shift, conducive to the consolidation environment. The significant reduction in the maximum principal stress, allows to affirm that the main deforming force, the shear, in this fracture pattern, was considerably reduced and the low value of Von Mises obtained, consistent with an implant capable of making an effective load sharing.