Long-term hearing performance and soft tissue outcomes of the Baha® Attract system in patients with bilateral congenital microtia in a single centre.
Ying ChenLiu-Jie RenNa GaoWen-Xiu GuChen-Long LiYao-Yao FuJun-Li ZhangYou-Zhou XieTian-Yu ZhangPublished in: Clinical otolaryngology : official journal of ENT-UK ; official journal of Netherlands Society for Oto-Rhino-Laryngology & Cervico-Facial Surgery (2021)
The Baha® Attract system can provide effective and stable hearing amplification for patients with microtia. Concurrent surgeries of auricle reconstruction and Baha® Attract implantation do not result in more soft tissue complications. The number of adverse events increased with the duration of implantation, even one year postoperatively. Benign skin evolutions, including skin thinning and erythema at the implant side, are commonly seen in Baha® Attract users. Implants are not necessarily removed when skin necrosis occurs in Baha® Attract users.