Detection of phosphorothioated (PS) oligonucleotides in horse plasma using a product ion (m/z 94.9362) derived from the PS moiety for doping control.
Teruaki TozakiKaoru KarasawaYohei MinamijimaHideaki IshiiMio KikuchiHironaga KakoiKei-Ichi HirotaKanichi KusanoShun-Ichi NagataPublished in: BMC research notes (2018)
We demonstrated the specific detection of phosphorothioated oligonucleotides (PSOs) using LC/MS/MS. PSOs produce the specific product ion (m/z 94.9362) derived from PS moiety. PS is not derived from endogenous substances in animal body, and the product ion is a suitable marker for the detection of PSOs. With our strategy, reproducible target analyses were achieved for identifying the specific substances, with a LOD of 0.1 ng/mL and a quantification rage of 0.1-200 ng/mL in deproteinated plasma. Non-target analyses could also detect the presence of PSOs selectively with 100 ng/mL in the same matrix. These results suggested that the detection of PSOs in horse blood is possible by targeting the product ion using LC/MS/MS.