Impact of magnetic field regulation in conjunction with the volumetric repainting technique on the spot positions and beam range in pencil beam scanning proton therapy.

Suresh RanaJaafar BennounaAlonso N GutierrezAnatoly B Rosenfeld
Published in: Journal of applied clinical medical physics (2020)
The volumetric repainting technique in magnetic field regulation mode resulted in acceptable spot position and range differences for our beam delivery system. The range differences were found to be within ±1 mm (TG224). For the spot positions (TG224: ±1 mm), the central axis measurements were within ±1 mm, whereas for the off-axis measurements, 97.9% of the spots were within ±1 mm, and all spots were within ±1.2 mm.