Login / Signup

Influence of patient head definition on induced E-fields during MR examination.

Tolga GorenSylvain RebouxSilvia FarcitoBryn LloydNiels Kuster
Published in: Magnetic resonance in medicine (2023)
The different head extents result in different maximum RF exposures affecting the risk assessment by up to a factor of 1.7. The results of this study can be used to estimate the additional uncertainty in safety assessments. Future revisions of MR standards should eliminate this inconsistency.