Login / Signup

Breathable Dry Silver/Silver Chloride Electronic Textile Electrodes for Electrodermal Activity Monitoring.

Peter A HaddadAmir ServatiSaeid SoltanianFrank KoPeyman Servati
Published in: Biosensors (2018)
The focus of this study is to design and integrate silver/silver chloride (Ag/AgCl) electronic textile (e-textile) electrodes into different textile substrates to evaluate their ability to monitor electrodermal activity (EDA). Ag/AgCl e-textiles were stitched into woven textiles of cotton, nylon, and polyester to function as EDA monitoring electrodes. EDA stimulus responses detected by dry e-textile electrodes at various locations on the hand were compared to the EDA signals collected by dry solid Ag/AgCl electrodes. 4-h EDA data with e-textile and clinically conventional rigid electrodes were compared in relation to skin surface temperature. The woven cotton textile substrate with e-textile electrodes (0.12 cm² surface area, 0.40 cm distance) was the optimal material to detect the EDA stimulus responses with the highest average Pearson correlation coefficient of 0.913 ± 0.041 when placed on the distal phalanx of the middle finger. In addition, differences with EDA waveforms recorded on various fingers were observed. Trends of long-term measurements showed that skin surface temperature affected EDA signals recorded by non-breathable electrodes more than when e-textile electrodes were used. The effective design criteria outlined for e-textile electrodes can promote the development of comfortable and unobtrusive EDA monitoring systems, which can help improve our knowledge of the human neurological system.
Keyphrases