Using XRF and ICP-OES in Biosorption Studies.
Katarzyna ChojnackaMateusz SamorajŁukasz TuhyIzabela MichalakMałgorzata MironiukMarcin MikulewiczPublished in: Molecules (Basel, Switzerland) (2018)
In this work, a method of recalculation of results of X-ray fluorescence (XRF) technique to Inductively Coupled Plasma-Optical Emission Spectroscopy (ICP-OES) method was elaborated for biosorption studies. Equations that calibrate XRF to ICP-OES were determined, as a biosorbent strawberry, blackcurrant and raspberry seeds after supercritical CO₂ extraction were used. ICP-OES showed a better precision and lower detection limits than XRF. The latter technique is cheaper, requires minimal sample preparation and gives faster results. Linear regression of the data gave almost 1:1 correlations without additional correction (for Cu r² = 0.9998, Mn r² = 0.807, Zn r² = 0.979). Calibration and quantification of intensities of XRF was obtained using ICP-OES measurements after samples digestion with HNO₃ in a microwave system. High positive correlations were estimated for Cu, Mn, Zn. It was demonstrated that XRF technique can be used together with other well established techniques (ICP-OES) to produce quantitative data from biosorption studies. Elaboration of cheap and quick analytical methodology is an important aspect in development of new processes and products based on biosorption process.