Cochlear Implant Device Failures Falling Under the 2020 FDA Voluntary Field Corrective Action: A Systematic Review and Meta-analysis.

Aparna GovindanMia SaadeEnrique R PerezGeorge B WannaMaura K CosettiZachary G Schwam
Published in: Otology & neurotology : official publication of the American Otological Society, American Neurotology Society [and] European Academy of Otology and Neurotology (2024)
The rate of pooled reported failure for CIs falling under the 2020 VFCA in the literature thus far is 23.7%. The overwhelming majority of these failures were device related, the rates of which were higher in children. Speech perception improved significantly after reimplantation.