Analysis of clinical data to determine the minimum number of sensors required for adequate skin temperature monitoring of superficial hyperthermia treatments.

Akke BakkerRebecca HolmanDario B RodriguesHana Dobšíček TrefnáPaul R StaufferGeertjan van TienhovenCoen R N RaschJohannes Crezee
Published in: International journal of hyperthermia : the official journal of European Society for Hyperthermic Oncology, North American Hyperthermia Group (2018)
Adequate coverage of the skin temperature distribution during superficial hyperthermia treatment requires the use of >50 stationary sensors per 400 cm2 applicator. Thermal mapping is a valid alternative.