Validation and reproducibility of cardiovascular 4D-flow MRI from two vendors using 2 × 2 parallel imaging acceleration in pulsatile flow phantom and in vivo with and without respiratory gating.

Jelena BockJohannes TögerSebastian BidhultKarin Markenroth BlochPer ArvidssonMikael KanskiHåkan ArhedenFrederik TestudAndreas GreiserEinar HeibergMarcus Carlsson
Published in: Acta radiologica (Stockholm, Sweden : 1987) (2018)
Sequence-specific in vivo validation of 4D-flow is needed before clinical use. 4D-flow with the Aera prototype sequence with a clinically acceptable acquisition time (<10 min) showed acceptable bias in healthy controls to be considered for clinical use. Intra-individual KE comparisons should use the same sequence.