Login / Signup

Absolute ion detection efficiencies of microchannel plates and funnel microchannel plates for multi-coincidence detection.

K FehreD TrojanowskajaJ GatzkeMaksim KunitskiF TrinterS ZellerL Ph H SchmidtJürgen StohnerR BergerA CzaschO JagutzkiT JahnkeR DörnerM S Schöffler
Published in: The Review of scientific instruments (2018)
Modern momentum imaging techniques allow for the investigation of complex molecules in the gas phase by detection of several fragment ions in coincidence. For these studies, it is of great importance that the single-particle detection efficiency ε is as high as possible, as the overall efficiency scales with εn, i.e., the power of the number of detected particles. Here we present measured absolute detection efficiencies for protons of several micro-channel plates (MCPs), including efficiency enhanced "funnel MCPs." Furthermore, the relative detection efficiency for two-, three-, four-, and five-body fragmentation of CHBrClF has been examined. The "funnel" MCPs exhibit an efficiency of approximately 90%, gaining a factor of 24 (as compared to "normal" MCPs) in the case of a five-fold ion coincidence detection.