Effects of different ablation settings on lesion dimensions in an ex vivo swine heart model: Baseline impedance, irrigant, and electrode configuration.

Xi JiangSijie LiQingsong XiongChangzhi ZhangLijuan PengWeijie ChenYangwei CaiYuehui YinShaojie ChenZhiyu Ling
Published in: Journal of cardiovascular electrophysiology (2022)
There is a strong negative linear correlation between baseline impedance and lesion dimensions at a certain AI value. Baseline impedance has an influence on the overall lesion dimensions among irrigated fluid and ablation configurations. Over a threshold impedance of 150 Ω, the predictive accuracy of AI can be compromised.