Cold Welding of the Synthes Femoral Neck System Encountered at Conversion to Arthroplasty: A Case Report.

Danielle F PetersonKathryn Schabel
Published in: JBJS case connector (2023)
Surgeons must recognize the need for additional available equipment (e.g., burr or broken screw removal set) for successful extraction when FNS removal is required.