Note: Development of real-time epithermal neutron detector for boron neutron capture therapy.

H TanakaY SakuraiT TakataTsubasa WatanabeShinji KawabataM SuzukiS-I MasunagaK TakiK AkaboriK WatanabeK Ono
Published in: The Review of scientific instruments (2018)
The real-time detection of epithermal neutrons forms an important aspect of boron neutron capture therapy. In this context, we developed an epithermal neutron detector based on the combination of a small Eu:LiCaAlF6 scintillator and a quartz fiber in order to fulfill the irradiation-field requirements for boron neutron capture therapy. The irradiation test is performed with the use of a reactor-based neutron source. The thermal and epithermal neutron sensitivities of our epithermal neutron detector are estimated to be 9.52 × 10-8 ± 1.59 × 10-8 cm2 and 1.20 × 10-6 cm2 ± 8.96 × 10-9 cm2, respectively. We also subject the developed epithermal neutron detector to actual irradiation fields, and we confirm that the epithermal neutron flux can be measured in realtime.