Biological variation of intact fibroblast growth factor 23 measured on a fully automated chemiluminescent platform.

Antonín JaborZdenek KubíčekJitka KomrskováTereza VackováJiří VymětalíkJanka Franeková
Published in: Annals of clinical biochemistry (2019)
The measurement of iFGF23 demonstrated a CVA less than 4% during the experimental estimation of biological variation. The total imprecision was less than 7% in the EP15-A2 experiment. The CVI values of iFGF23 in healthy persons were 14.2 (SD ANOVA) and 12.5% (CV ANOVA), respectively. The CVG was 13.4%, and the resulting index of individuality was 1.06. The reference change value was less than 41%. The availability of this automated assay for iFGF23 with well-characterized biological variation data delivers opportunities for improved availability and application of this assay clinically.