Despite the use of a common WHO reference material for total IgE assay calibration, significant differences in quantitation was observed between two FDA-cleared test systems. Given that, it is warranted for clinical laboratories to verify vendor established reference intervals and adjust accordingly based on internal assessment of the normal range.