Improving the robustness of pseudo-continuous arterial spin labeling to off-resonance and pulsatile flow velocity.
Li ZhaoMarta VidorretaSalil SomanJohn A DetreDavid C AlsopPublished in: Magnetic resonance in medicine (2016)
The robustness of PCASL can be increased by selecting an unbalanced scheme with a low average gradient (0.5 mT/m), a low ratio (7×) of selective to average gradients, and the highest feasible B1 (1.8 μT). Placing the labeling plane above the carotid bifurcation and below the V3 segment, usually between the second and third vertebrae, yielded robust results. Magn Reson Med 78:1342-1351, 2017. © 2016 International Society for Magnetic Resonance in Medicine.