Optical crosstalk of protective cover on MPPC array for TOF PET detector.
Eiji YoshidaFujino ObataTaiga YamayaPublished in: Physics in medicine and biology (2024)
powder as the reflective material. 

Main results. 
Coincidence counts were measured with a pair of TOF detectors, and the CRT was shorter with a deeper slit depth. The CRT before improvement was 235 ps, and using the cover having the 350 μm deep slits filled with reflective material lowered the CRT to 211 ps. 

Significance. 
Up to 10% of the scintillation light was diffused to other MPPCs by the protective cover, and the CRT was degraded by 10% due to optical crosstalk of the cover. The proposed method promises to improve the CRT of the TOF detector.