Impact of air gap, range shifter, and delivery technique on skin dose in proton therapy.

A KernChristian BäumerK KröningerJörg WulffB Timmermann
Published in: Medical physics (2020)
The study investigated aspects of skin dose of a water equivalent phantom by varying key parameters of a proton treatment plan. Parameters like the RS, the air gap, and the delivery modality have an impact on the order of 4.0% for the skin dose at the depth of 70 μm. The increases in skin dose are the effects of the contribution of the increased electron fluence at small air gaps and the emitted hadronic particles produced by the RS.