Login / Signup

Entrapment of diagnostic catheter within Advisor HD grid mapping catheter.

Philip L MarLuke ChongArturo PerezDhanunjaya R LakkireddyRakesh Gopinathan Nair
Published in: Journal of cardiovascular electrophysiology (2021)
The potential for catheter entanglement with the HD Grid mapping catheter is explicitly stated in the manufacturer's product manual. A case of an entrapped 6 French quadripolar diagnostic catheter within an HD Grid mapping catheter is presented. We discuss the diagnosis, management, and resolution of this complication in our patient. The patient's arrhythmia was successfully eliminated, and no vascular complication in the postprocedural setting nor arrhythmia recurrence at follow-up were observed. Strategies to prevent and safely manage this complication, while salvaging access, are also discussed.
Keyphrases
  • ultrasound guided
  • high resolution
  • case report
  • high density
  • heart failure
  • mass spectrometry