Disregarded Measurement Uncertainty Contributions and Their Magnitude in Measuring Plasma Glucose.
Astrid PetersmannRainer MacdonaldMatthias NauckPublished in: Journal of diabetes science and technology (2020)
Estimation of MU in this study is quite reliable due to the large number of IQC data from one year. The MU of the target values of the commercial control material in this study was considerably larger than other MU contributions, ie, standard deviation and bias. In the future, the contribution of MU of commercial IQC should be addressed more carefully and technologies to measure glucose should be geared toward smaller MU possible, as needed, especially for glucose concentration measurements in diagnosis and management of DM.