Login / Signup

Evaluation of a Refined Implantable Resonator for Deep-Tissue EPR Oximetry in the Clinic.

Eunice Y ChenDan TseHuagang HouWilson A SchreiberPhilip E SchanerMaciej M KmiecKendra A HebertPeriannan KuppusamyHarold M SwartzBenjamin B Williams
Published in: Applied magnetic resonance (2021)
IRs were functional for up to 9 months after implantation and provided deep tissue oxygen measurements using EPR oximetry. Tissues surrounding the IRs showed no more tissue reaction than tissues surrounding the control samples. This pre-clinical study demonstrates that the IRs can be safely implanted in brain and leg tissues and that repeated, non-invasive, deep-tissue oxygen measurements can be obtained using in vivo EPR oximetry.
Keyphrases
  • gene expression
  • white matter
  • clinical trial
  • brain injury
  • subarachnoid hemorrhage