Analytical verification of the Atellica VTLi point of care high sensitivity troponin I assay.

Christopher M FlorkowskiVanessa BuchanBobby Vincent LiFelicity TaylorMinh PhanMartin P ThanJohn William Pickering
Published in: Clinical chemistry and laboratory medicine (2024)
Concordances between VTLi and laboratory assays were at least non-inferior to those between laboratory assays. Imprecision met manufacturer claims and was consistent with a high sensitivity assay. There is potential for hemolysis interference, highlighting the need for quality samples. The results support performance characteristics previously reported in validation studies, and the device offers acceptable performance for use within intended medical settings.