Comparison of the performance of two real-time fluorescent quantitative PCR kits for the detection of SARS-CoV-2 nucleic acid: a study based on large real clinical samples.

Yiting WangXuewen LiYifei WangZheyu TuJian-Cheng XuJunqi PanQi Zhou
Published in: Virology journal (2022)
Since the Maccura kit is found to have false positives in actual clinical work, recalculation of the cut-off values can reduce this occurrence. In order to improve the accuracy of the testing, laboratories should use two kits for COVID-19 testing, and the adjusting and optimizing of the kits for their situation are needed.