Analysis of clinical patient-specific pre-treatment quality assurance with the new helical tomotherapy platform, following the AAPM TG-218 report.
Marco FusellaSamuele CavinatoAlessandra GermaniMarta PaiuscoNicola PivatoMarco Andrea RossatoAnthony ScottAlessandro ScaggionPublished in: Radiation oncology (London, England) (2021)
The phantom measurements closely match the planned dose distributions, demonstrating that the calculation accuracy of the new Precision TPS and the delivery accuracy of the Radixact unit are adequate, with respect to international guidelines and reports. Furthermore, a first correlation with the planning parameters was made. Action and tolerance limits have been set for the new Radixact Linac.