Towards non-invasive tissue hydration measurements with optical coherence tomography.
Linda B NeubrandTon G van LeeuwenDirk J FaberPublished in: Journal of biophotonics (2024)
The attenuation coefficient ( μ OCT ) measured by optical coherence tomography (OCT) has been used to determine tissue hydration. Previous dual-wavelength OCT systems could not attain the needed precision, which we attribute to the absence of wavelength-dependent scattering of tissue in the underlying model. Assuming that scattering can be described using two parameters, we propose a triple/quadrupole-OCT system to achieve clinically relevant precision in water volume fraction. In this study, we conduct a quantitative analysis to determine the necessary precision of μ OCT measurements and compare it with numerical simulation. Our findings emphasize that achieving a clinically relevant assessment of a 2% water fraction requires determining the attenuation coefficient with a remarkable precision of 0.01 m m - 1 . This precision threshold is influenced by the chosen wavelength for attenuation measurement and can be enhanced through the inclusion of a fourth wavelength range.