Two-Dimensional Shear Wave Elastography versus Transient Elastography: A Non-Invasive Comparison for the Assessment of Liver Fibrosis in Patients with Chronic Hepatitis C.
Rosanna VillaniFrancesco CavalloneAntonino Davide RomanoFrancesco BellantiGaetano ServiddioPublished in: Diagnostics (Basel, Switzerland) (2020)
In recent years, several non-invasive methods have been developed for staging liver fibrosis in patients with chronic hepatitis C. A 2D-Shear wave elastography (SWE) technique has been recently introduced on the EPIQ 7 US system (ElastQ), but its accuracy has not been validated in patients with chronic hepatitis C virus (HCV) infection. We enrolled 178 HCV patients to assess their liver fibrosis stage with ElastQ software using transient elastography as a reference standard. The best cut-off values to diagnose ³ F2, ³ F3, and F4 were 8.15, 10.31, and 12.65 KPa, respectively. Liver stiffness values had a positive correlation with transient elastography (r = 0.57; p < 0.001). The area under the receiver operating characteristics (AUROC) was 0.899 for ³ F2 (moderate fibrosis), 0.900 for ³ F3 (severe fibrosis), and 0.899 for cirrhosis. 2D-SWE has excellent accuracy in assessing liver fibrosis in patients with chronic hepatitis C and an excellent correlation with transient elastography.