Login / Signup

Validation of a novel automated signal analysis tool for ablation of Wolff-Parkinson-White Syndrome.

Scott R CeresnakRobert H PassAnne M DubinLingyao YangKara S MotonagaHaley HedlinKishor AvasaralaAnthony V TrelaDoff B McElhinneyChristopher JansonLynn NappoXuefeng B LingGregory J Gates
Published in: PloS one (2019)
An automated signal analysis software tool reliably distinguished successful versus unsuccessful ablation electrograms in children with WPW when validated in a large, diverse cohort. Refining the tools using an alternative threshold and statistical method did not improve the original signal score at a threshold of 3.1. This software was effective across two centers and multiple operators and may be an effective tool for ablation of WPW.
Keyphrases
  • young adults
  • deep learning
  • case report