Login / Signup

Trace Oxygen Affects Osmium Redox Polymer Synthesis for Wired Enzymatic Biosensors.

Margaret C CalhounChristopher D StachurskiSara L WinnEvan A GizzieAaron W DanielNathan D SchleyDavid E Cliffel
Published in: Journal of the Electrochemical Society (2022)
Electrochemical sensors that utilize enzymes are a sensitive, inexpensive means of detecting biologically relevant analytes. These sensors are categorized based on their construction and method of signal transport. Type I sensors consist of a crosslinked enzyme on an electrode surface and are potentially subject to interference from byproducts and other biological analytes. However, type II sensors help alleviate this problem with the addition of a redox polymer layer that assists in signal transduction, thus minimizing interferences. An osmium-loaded poly(vinylimidazole) polymer (Os-PVI) is commonly used with successful results, and when combined with an enzyme yields a type II sensor. Our initial attempts at the synthesis of this polymer resulted in an unexpected osmium precursor, which had fluorescent and redox properties that did not match with the desired Os-PVI polymer. Careful exclusion of oxygen during the Os complex precursor synthesis was necessary to avoid this unexpected oxygen containing Os-precursor, which had been seen previously in mass spectrometry studies. All precursors and osmium polymers were characterized with 1 H NMR, fluorescence, mass spectrometry, and cyclic voltammetry to provide a better understanding of these compounds and assist in the building of new sensors.
Keyphrases