Login / Signup

A Europium Nanosphere-Based Time-Resolved Fluorescent Immunochromatographic Assay for the Rapid Screening of 4,4'-Dinitrocarbanilide: Aiming at Improving Strip Method Performance.

Ming ZouYongkang YinLiuchuan GuoQidi ZhangJinyan LiHong ZhangQian SongZhaojie LiLi WangXiang AoXiao Liang
Published in: Biosensors (2023)
Considering that the strip method is simple and convenient for users, a Europium nanosphere-based time-resolved fluorescent immunochromatographic assay (TRFICA) for the rapid screening of 4,4'-dinitrocarbanilide (DNC) was developed to improve the performance of strip assays. After optimization, TRFICA showed IC 50 , the limit of detection, and cut-off values of 0.4, 0.07, and 5.0 ng mL -1 , respectively. No significant cross-reactivity (CR < 0.1%) with 15 DNC analogs was observed in the developed method. TRFICA was validated for DNC detection in spiked chicken homogenates, and recoveries ranged from 77.3% to 92.7%, with coefficients of variation of <14.9%. Moreover, the time needed for the detection procedure, including the sample pre-treatment, was less than 30 min for TRFICA, which had never been achieved before in other immunoassays. The newly developed strip test is a rapid, sensitive, quantitative, and cost-effective on-site screening technique for DNC analysis in chicken muscle.
Keyphrases
  • loop mediated isothermal amplification
  • label free
  • sensitive detection
  • high throughput
  • quantum dots
  • real time pcr
  • skeletal muscle
  • molecular dynamics simulations