A point-of-care aptasensor based on the upconversion nanoparticles/MoS 2 FRET system for the detection of Pseudomonas aeruginosa infection.

Dou CheXitao CaoChong ChenHui Yan
Published in: Mikrochimica acta (2023)
The rapid detection of Pseudomonas aeruginosa (P. aeruginosa) is of great significance for the diagnosis of medical infection. In view of the above, a novel aptasensor based on fluorescence resonance energy transfer (FRET) was developed. It contained aptamer-coupled upconversion nanoparticles (UCNPs-apt) as a donor (excitation 980 nm) and molybdenum disulfide (MoS 2 ) nanosheets as an acceptor. The upconversion fluorescence aptamer system was investigated to obtain the optimal parameters of MoS 2 concentration, the incubation time of UCNPs-apt/MoS 2 and P. aeruginosa, and pH. Based on the optimal parameters, a linear calibration equation (emission 654 nm) with a wide detection range 8.7 × 10 ~ 8.7 × 10 7  cfu/mL, a high coefficient of determination R 2 0.9941, and a low limit of determination (LOD) 15.5 cfu/mL were established. The method was validated with P. aeruginosa infected foci of mouse wound. The advantage of this aptasensor is that analysis results can be obtained  within 1.5 h, which was much faster than that of the standard method (18-24 h). Furthermore, combined with a portable instrument, it can be used as a point-of-care testing for the early detection of P. aeruginosa infection, which is useful for selecting the correct antibiotics to achieve good therapeutic effects. Additionally, it also has a broad application prospect in food and environmental areas.