Login / Signup

Facial nerve stimulation in cochlear implant users - a matter of stimulus parameters?

Lutz GärtnerThomas LenarzJurgita IvanauskaiteAndreas Büchner
Published in: Cochlear implants international (2022)
Aberrant facial nerve stimulation in cochlear implant patients was successfully resolved by re-implantation. We hypothesize that one or more of the specific stimulus paramet ers are responsible for this success: combined common ground and monopolar stimulation mode, asymmetric pulse waveform with active anodic leading first phase followed by slow capacitive discharge, and pulse width modulation for loudness coding.